Oro / Oro Plus

GE GlucoCapture

Oro / Oro Plus

Never Lose Track Again

The GM700S is a premium meter featuring FAD-GDH reagent technology to minimize oxygen interference and keep you updated with results you can trust. Accepts a wide HCT range and comes with helpful meal markers to keep you as informed as possible.

Full compliance with EN ISO 15197: 2015 and ISO 15197: 2013

CEFDAISO13485_2016IVD

pro-img/GE_product/oro.jpg
pro-img/GE_product/oro_plus.jpg
pro-img/04 GM700SB/En/03_GS700.jpg
  • Wide HCT range

    Auto-Coding

    Eliminates the need for manual coding and delivers maximum accuracy.

  • Meal markers

    Wide HCT Range

    Suitable for neonatal and anemic patients.

  • Multiple blood sample

    Various Blood Samples Accepted

    Allows capillary, venous and arterial blood.

  • Bluetooth transmission

    Complete Blood Sugar Management*

    Add notes, photos and status indicators for more complete entries.

  • Multiple blood sample

    Personalization*

    Set blood glucose targets and customize your measurement schedule.

  • Auto coding

    Intuitive Statistics Dashboard*

    Useful data presented clearly.

*Bluetooth connectivity available with GM700SB only. Transfer results results wirelessly to the Rightest CARE App.

  • 1
    Insert the strip into the socket of the meter. Calibration will occur automatically.
  • 2
    Touch a blood sample against the sample entry port of the strip.
  • 3
    After 5 seconds, the test result will appear.
  • 4
    Rotate and pull upwards to remove the test strip.
Test strip GS700
Coding Auto-coding
Enzyme / Measurement technology GDH / Electrochemical sensor
Blood sample source Capillary, venous, arterial and neonatal whole blood
Minimum sample volume 0.75 µL
HCT range Glucose concentrations ≦ 11.1mmol/L(200mg/dL), HCT range 20 - 70%
Glucose concentrations > 11.1mmol/L(200mg/dL), HCT range 20 - 60%
Measuring range 0.6 - 33.3 mmol/L (10 - 600 mg/dL)
Meter dimensions (L x W x H) 82.0 x 45.0 x 15.5 mm (3.22 x 1.77 x 0.61 inch)
Display area (L x W) 34.0 x 27.5 mm (1.09 x 1.34 inch)
Weight (with battery) 43.0 ± 5 g
Power supply 1 × CR2032 battery
Memory capacity 500 blood glucose test results with date and time
Operating temperature 6 - 44℃ (43 - 111°F)
Operating relative humidity 10 - 90%
Meter storage conditions -10 - 60°C (14 - 140°F)
Test strip storage conditions 4 - 30°C (39 - 86°F), < 90% relative humidity
Bluetooth version 4.0
Accuracy 95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%
Precision Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%

Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)