GE GlucoCapture Oro Series

GE GlucoCapture

Oro Series

Stay connected. Stay in control.

Never lose track again. The GM700SB features Bluetooth connectivity to let you manage your glucose straight from your phone with the Rightest CARE App. A wide HCT-range, acceptance of a variety of blood samples and helpful meal markers feature deliver an informative, user-friendly experience.
ISO15197_2013CEFDAISO13485_2016IVD

pro-img/04 GM700SB/En/01_GM700SB_mmol.jpg
pro-img/04 GM700SB/En/02_GM700SB_mmol.jpg
pro-img/04 GM700SB/En/03_GS700.jpg
pro-img/04 GM700SB/En/04_GD500.jpg
  • Wide HCT range

    Bluetooth Connectivity

    Transfer test results to smartphones wirelessly.

  • Meal markers

    Wide HCT range

    Measures glucose values for extreme hematocrit ranges including neonatal, elderly and anemia patients.

  • Multiple blood sample

    Meal markers

    Before-and-after meal markers help you understand how food influences your blood glucose.

  • Auto coding

    Multiple Test Sites

    Allows capillary, venous, arterial and neonatal whole blood samples.

  • Proprietary Rightest® strip

    Auto-Coding

    Eliminates the needs for manual coding and ensures high accuracy.

  • Blood contact prevention

    Proprietary Rightest strip

    With Validus Technology™ strip design, Rightest™ strips ensure high accuracy and precision in readings.

GE GlucoCapture Oro Meter

With Rightest CARE App

Glucose Management Made Simple

Rightest CARE App QR Code
  • 1
    Insert the strip in the socket of the meter, which will do calibration automatically.
  • 2
    Place blood sample against the sample entry of strip.
  • 3
    After 5 seconds, the test result appears.
  • 4
    Rotate and pull up the test strip.
Test strip GS700
Coding Auto coding calibration
Enzyme / Measurement technology GDH / Electrochemical sensor
Blood sample source Capillary, venous, arterial and neonatal whole blood
Minimum sample volume 0.75 microliters (0.00075c.c.)
HCT range Glucose concentrations ≦ 11.1mmol/L(200mg/dL), HCT range 20 - 70%
Glucose concentrations > 11.1mmol/L(200mg/dL), HCT range 20 - 60%
Measuring range 0.6 - 33.3 mmol/L (10 - 600mg/dL)
Meter dimensions (L x W x H) 82.0 x 45.0 x 15.5 mm (3.22 x 1.77 x 0.61 inch)
Display area (L x W) 34.0 x 27.5 mm (1.09 x 1.34 inch)
Weight (with battery) 43.0 ± 5 g
Power supply 1 CR2032 battery
Memory capacity 500 blood glucose test results with date and time
Operating temperature 6 - 44℃ (43 - 111°F)
Operating relative humidity 10 - 90%
Meter storage conditions -10 - 60°C (14 - 140°F)
Test strip storage conditions 4 - 30°C (39 - 86°F), < 90% relative humidity
Bluetooth version 4.0
Accuracy 95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%
Precision Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%

Fully compliant with ISO 15197:2013 (For other received credentials, please contact us.)