ALPHA

Rightest Glucose Meter

ALPHA

Designed with Your Needs in Mind

The Alpha comes fitted with advanced features including auto-coding and a backlit display to provide quality, reliable blood glucose management.

Full compliance with EN ISO 15197: 2015 and ISO 15197: 2013

CEISO13485_2016IVD

pro-img/ALPHA/ALPHA_meter.jpg
pro-img/ALPHA/ALPHA_meter+strip.jpg
  • Fast Results

    Fast Results

    Displayed in 5 seconds.

  • Large Display

    Large Display

    Easy to view results.

  • Side-Insert Design

    Side-Insert Design

    Easy to handle and prevents blood contact for excellent hygiene.

  • Proprietary Strips

    Proprietary Strips

    Enhanced size for improved grip, ensuring accuracy and precision.

  • User-Friendly

    User-Friendly

    Start monitoring quickly with this simple design.

  • 1
    Insert the strip into the socket of the meter. Calibration will occur automatically.
  • 2
    Touch a blood sample against the sample entry port of the strip.
  • 3
    After 5 seconds, the test result will appear.
  • 4
    Rotate and pull upwards to remove the test strip.
Test strip ALPHA strip
Coding Coding not required
Enzyme / Measurement technology GOD / Electrochemical sensor
Blood sample source Capillary whole blood
Minimum sample volume 0.75 µL
HCT range 30 - 57%
Measuring range 0.6 - 33.3 mmol/L (10 - 600mg/dL)
Meter dimensions (L x W x H) 90.6 x 46.0 x 16.5 mm (3.56 x 1.81 x 0.64 inch)
Display area (L x W) 46.6 x 33 mm (1.83 x 1.32 inch)
Weight (with battery) 53.0 ± 5 g
Power supply 1 × CR2032 battery
Memory capacity 500 test results with date and time
Operating temperature 10 - 40°C (50 - 104°F)
Operating relative humidity 10 - 90%
Meter storage conditions -10 - 60°C (14 - 140°F)
Test strip storage conditions At 4 - 30°C (39 - 86°F), and below 90% relative humidity
Accuracy For glucose concentrations < 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements.
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 15% of reference measurements.
Precision For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%

Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)