GT200

Rightest glucose meter GT200 

ISO15197_2013fda_wiz-plus_rightestISO13485_2016IVD

pro-img/GT200/en/01_GT200_mmol.jpg
pro-img/GT200/en/02_GT200_mmol.jpg
pro-img/GT200/en/03_GT200.jpg
pro-img/GT200/en/04_GD500.jpg
  • proimages/pro-icons/website_icon-05.png

    Auto coding

    Code embossed on every strip for increased accuracy.

  • proimages/pro-icons/website_icon-09.png

    Large backlight screen

    Worry-free to use in dark environment.

  • proimages/pro-icons/website_icon-02.png

    Innovation strip design

    Larger, thicker, test strip help with dexterity issues.The innovative side - insert design ensures better hygiene. 

  • proimages/pro-icons/website_icon-10.png

    Wide HCT range

    Measures glucose value of extreme hematocrit range, like neonate, elder, anemia patient.

  • proimages/pro-icons/website_icon-13.png

    Diverse blood sample

    Tests capillary, venous, arterial and neonatal whole blood.

  • proimages/pro-icons/website_icon-46.png

    Reliable accuracy

    Full compliance ISO 15197:2013.

  • 1 proimages/pro-step/24_GT200/01_GT200-Step.jpg
    Friendly - large strip hand bar
  • 2 proimages/pro-step/24_GT200/02_GT200-Step.jpg
    Place blood sample against the sample entry of strip.
  • 3 proimages/pro-step/24_GT200/03_GT200-Step.jpg
    Rapidly - results in 5 seconds
  • 4 proimages/pro-step/24_GT200/04_GT200-Step_en.jpg
    Safety - keep from touching blood sample
Test strip GT200
Coding Auto coding
Enzyme / Measurement technology GOD / Electrochemical sensor
Blood sample source Capillary, venous, arterial and neonatal whole blood
Minimum sample volume 0.75 microliters  (0.00075c.c.)
HCT range Glucose concentrations ≦ 11.1mmol/L(200mg/dL), HCT range 20 - 70%
Glucose concentrations > 11.1mmol/L(200mg/dL), HCT range 20 - 60%
Measuring range 0.6 - 33.3 mmol/L (10 - 600mg/dL)
Meter dimensions (L x W x H) 96.0 x 46.0 x 17.5 mm
Display area (L x W) 55.0 x 37.0 mm
Weight (with battery) 65.0 g ± 5 g
Power supply 2 CR2032 batteries
Memory capacity 1,000 blood glucose test results with date and time
Operating temperature 10 - 40°C (50 - 104°F)
Operating relative humidity 10 - 90%
Meter storage conditions -10 -60°C (14 - 140°F)
Test strip storage conditions 4 - 30°C (39 - 86°F), < 90% relative humidity
Accuracy 95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%
Precision Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%

Fully comply with ISO 15197:2013 (For other received credentials, please contact us.)

proimages/pro-download/GT200/GT200_mmol.png
  • Instructions for use
  • Test Strip Guide
  • Instruction for Lancing Device